| Demand of power supply | Capacity | Two-phase 30KVA |
| Voltage frequency | 380V / 220V10% 50HZ ±0.5HZ | |
| Internal resistance | 0.75 at 380V; 0.25 at 220V | |
| Power supply | Adjustable range | 380V±10% at 380V; 220V±10% at |
| Photograph | Voltage | 50~100KV, *50~125KV being adjustable continuously |
| Current | small focus: 50mA, 100mA big focus: 50mA, 100mA, 200mA, 300mA | |
| Time | 0.04~6.3s, 23 grades together grading according to R10 coefficient | |
| High-voltage generator | Capacity | 20KV A(instantaneous) |
| Max. DC output voltage | 100KV 125KV | |
| Max. DC output current | 200mA 300mA | |
| X-Ray tube | Model | XD51-20, 40/100 XD51-20, 40/125 |
| Focus | small focus: 1x1mm; big focus: 2x2mm | |
| Photographic bed(YZ-SCI) | Length x Width x Height of bed face | 2000x710x700mm |
| Moving range of bed face | 680mm in length; 200mm in breadth | |
| Ray-filter of bed | Travel in length≥500mm | |
| Grid density N28 | ||
| Convergence distance fo100cm | ||
| Grid ratio r8 | ||
| Pillar of X-Ray tube unit | Moving in length along photographic bed | 1800mm or 1600mm |
| Distance to ground as moving up and down (along upright post) | 650~1800mm | |
| Rotary around the center or cross arm | ±180° | |
| Rotary around the shaft line of X-Ray annular tubes | -10° ~60° ~+120° | |
| Max size of cassette for photograph | 356x432mm(14" x 17") |